Purpose :-
Babies who are born too early, before 37 weeks, can have problems such as low birth weight, irregular temperature, and unstable vital signs. A baby incubator helps control their temperature. They will also be given high-calorie formula and will get the treatment they need for any other issues
Alarming:
Audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater (except for the low deviation alarm), meanwhile, the skin temperature indication window can indicate the alarm code. For the alarm 1-6, the relevant light is on, and for the alarm 7, all alarming lights are on; All alarm sound can be cancelled except for the power failure alarm and the system alarm.
Over-temp alarm: When the temperature ≤ 38º C;
Upper Deviation alarm: Air temperature control 3.0º C
Lower Deviation alarm: Air temperature control -3.0º C
Sensor alarm: When the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: When the fan is blocked, speed ≤ 1000rpm or stop working;
Power failure alarm: When there is no power or the power cord is disconnected.
System alarm: When there is E2ROM malfunction inside of temperature controller.
Air mode control, set temperature and air temperature are displayed separately;
Natural air flow humidity;
Failure alarm indication;
RS-232 connector.